Finding Description
2. During the initial observation, on 3/6/23 at 2:35 p.m., Resident 16 was lying in his bed with his supplemental oxygen (O2) being administered via nasal canula (NC-a medical device to provide supplemental oxygen therapy to people who have lower oxygen levels). The resident's O2 concentrator (a medical device that provides supplemental oxygen) was observed to be dust covered. The humidification bottle (a sealed bottle of water inserted into a breathing circuit to add moisture to the breathing gases for administration) was not dated. A nebulizer (an electrically powered machine that turns liquid medication into a mist so that it can be breathed directly into the lungs through a face mask or mouthpiece) and tubing were observed in a bag which was dated 2/23/23.
During a random observation, on 3/7/23 at 2:18 p.m., the resident was lying in his bed with his NC in place. The O2 concentrator was observed to be set to administer his O2 at 3 liters (L). The resident's O2 concentrator was observed to be dust covered. The humidification bottle was not dated. A O2 tubing bag, attached to the concentrator, was dated 2/23/23. The nebulizer mask was observed in a bag dated 2/23/23, and noted to have brown debris inside of the mask, near where it would be placed on the resident's mouth.
During a random observation of the resident's room, on 3/8/23 at 10:55 a.m., the resident was not in the room. The O2 concentrator was turned off. The O2 concentrator was observed to be dust covered. The nebulizer mask appeared to have been changed to a new mask. No debris noted on the mask. No date noted on the bag which contained the nebulizer. The humidification bottle on the O2 concentrator was not dated.
During a random observation, of the resident's room, on 3/8/23 at 11:56 a.m., the resident was not in the room. The O2 concentrator was turned off. The O2 concentrator was observed to be dust covered. The O2 NC tubing was in a bag, dated 2/23/23, hanging on the back of the concentrator. The tubing was observed to have clear fluid drops in the tubing. None of the clear fluid was noted on the inside of the bag where the tubing was placed.
During a random observation on, 3/8/23 at 12:00 p.m., the resident was sitting in the dining room. The resident's supplemental O2 was being administered through a portable O2 tank. The O2 tank was set at 3 L and was being administered via NC. The NC was noted to be out of the nostrils to the left side of his nose.
During a random observation, on 3/9/23 at 10:15 a.m., the resident was lying in bed with his O2 NC in place. The nebulizer tubing and mask were in an undated bag. The inside of nebulizer mask was observed to be soiled with a dust like debris and oily residue. The tubing, mask, and nebulizer chamber were dry. At the same time, the resident indicated the nurse puts the mask on and takes it off for him.
Resident 16's record was reviewed on 3/8/23 at 1:50 p.m. The profile indicated the resident's diagnoses included, but were not limited to, chronic obstructive pulmonary disease (COPD-a group of diseases that cause airflow blockage and breathing-related problems).
An annual Minimum Data Set (MDS-part of the federally mandated process for clinical assessment of all residents in Medicare and Medicaid certified nursing homes) assessment, dated 11/11/22, indicated the resident had severe cognitive deficit, was short of breath (SOB) when lying flat, and required supplemental O2.
A care plan, dated 10/13/21, indicated the resident required supplemental O2 due to his diagnosis of COPD. Interventions included, but were not limited to, administer O2 per physician's order at 2 L per NC.
A physician's order, dated 8/24/21, indicated change oxygen tubing and humidity. Clean concentrator and filter. Once A Day on Sunday.
A physician's order, dated 10/13/21, indicated O2 at 2 L per NC continuously.
A physician's order, dated 10/13/21, indicated change oxygen tubing and water canister (humidification bottle) every week on Sunday night shift. Label tubing with date and initials.
A physician's order, dated 1/7/23, indicated ipratropium-albuterol solution (a medication used to help control the symptoms of lung diseases, such as asthma, chronic bronchitis, and emphysema) for nebulization; 0.5 milligrams (mg)-3 mg (2.5 mg base)/3 milliliters (mL). 1 vial inhalation two times a day.
3. During the initial observation, on 3/6/23 at 10:24 a.m., Resident 1 was lying in bed with supplemental oxygen (O2) being administered by nasal canula (NC-a medical device to provide supplemental oxygen therapy to people who have lower oxygen levels). The resident's O2 concentrator (a medical device that provides supplemental oxygen) was set at 2.5 liters (L). The date on O2 tubing bag was 2/23/23, and the date on the humidification bottle (a sealed bottle of water inserted into a breathing circuit to add moisture to the breathing gases for administration) was 3/5/23.
During a random observation, on 3/7/23 at 10:00 a.m., the resident was lying in bed with O2 being administered at 2.5 L by NC. The date on O2 tubing bag was 2/23/23, and the date on the humidification bottle was 3/5/23.
During a random observation, on 3/8/23 at 10:41 a.m., the resident was lying in bed with O2 being administered at 2.5 L by NC. The date on O2 tubing bag was 2/23/23, and the date on the humidification bottle was 3/5/23.
During a random observation, on 3/9/23 at 10:00 a.m., the resident was lying in bed with O2 being administered at 2.5 L by NC. The date on O2 tubing bag was 2/23/23, and the date on the humidification bottle was 3/5/23.
Resident 1's record was reviewed on 3/9/23 at 1:15 p.m. The profile indicated the resident's diagnoses included, but were not limited to, chronic obstructive pulmonary disease (COPD-a group of diseases that cause airflow blockage and breathing-related problems) and congestive heart failure (CHF-occurs when the heart muscle doesn't pump blood as well as it should).
A quarterly Minimum Data Set (MDS-part of the federally mandated process for clinical assessment of all residents in Medicare and Medicaid certified nursing homes) assessment, dated 12/9/22, indicated the resident had moderate cognitive deficit, was short of breath (SOB) when lying flat, and required supplemental O2.
A care plan, dated 10/13/21, indicated the resident required O2 per NC continuously related to his diagnosis of CHF and COPD. Interventions included, but were not limited to, administer O2 per physician's order order and staff to change O2 tubing and water canister (humidification bottle) every week on Sunday night shift. Label tubing with date and initials.
A physician's order, dated 2/17/21, indicated O2 at 3 liters (L) per NC continuously.
A physician's order, dated 12/10/22, indicated change O2 tubing and humidification bottle, clean concentrator and filter, once daily on Sunday.
During an interview, on 3/10/23 at 11:57 a.m., Licensed Practical Nurse (LPN) 6 indicated, there were 4 or 5 residents who had orders for oxygen, and 2 who received nebulizer treatments. The residents must have a physician's order for oxygen use and for nebulizer treatments. Nebulizer hand piece and oxygen masks were replaced with new units weekly, and placed back into storage bags. Both oxygen and nebulizer equipment had physician's orders to be changed weekly on Sunday night shift, and to be documented in the medication administration record (MAR) in electronic medical record (EMR). Equipment should be dated when changed. If equipment was soiled in any way in between scheduled change dates the nurse would be responsible for cleaning it or changing the soiled piece of equipment. Cleaning of the nebulizer or oxygen concentrators should be cleaning at the same time the tubing, masks and bottles, to include the filter as needed.
During an interview, on 3/10/23 at 12:05 p.m., LPN 6 indicated, the process for nebulizer treatments were, there must be a physician's order for medication. The resident's vital signs (respirations, oxygen saturations, listen to breath/lung sounds) should be completed. The medication would then be placed into the nebulizer hand piece. Hand the nebulizer to resident to hold. Mid way through the procedure, recheck vitals and breath sounds. When finished help resident place nebulizer in bag, and recheck vitals.
On 3/10/23 at 1:50 p.m., the Executive Director (ED) provided an undated document, titled, Oxygen Therapy and Devices, and indicated it was the policy currently being used by the facility. The policy indicated, .Definition of Oxygen: 1) Oxygen is a drug which must be ordered by a physician .Initiation of Oxygen: 1) Verify physician order .7) Apply device to the patient with appropriate liter flow .Oxygen Devices: 1) Nasal Canula .e. Change out weekly and PRN (as needed). f. Place in a labeled bag when not in use
3.1-47(a)(6)
Based on observation, interview, and record review, the facility failed to properly clean and store nebulizer (small machine that turns liquid medication into a mist that can be easily inhaled) and oxygen equipment for 3 of 4 residents reviewed for respiratory care (Residents 21, 1, and 16).
Findings include:
1a. On 3/6/2023 at 1:53 p.m., Resident 21's nebulizer tubing and equipment was observed to be in a clear plastic bag and was unlabeled and not dated. The clear plastic bag was hanging from the nebulizer machine.
On 3/7/2023 at 11:01 a.m., Resident 21's nebulizer tubing and equipment including mouthpiece was observed to be out of the bag sitting next to the nebulizer machine. No date noted to be on the tubing. Resident was resting in her recliner resting. The clear plastic bag for nebulizer observed to be on the floor next to side table.
On 3/7/2023 at 12:00 p.m., Resident 21's nebulizer tubing and equipment including mouthpiece was observed to be out of the bag sitting next to machine. Resident was resting in her recliner.
On 3/7/2023 at 12:59 p.m., Resident 21's nebulizer tubing and equipment including mouthpiece was observed to be out of the bag sitting next to machine. Resident was sitting up in wheelchair eating lunch. Clear plastic bag for nebulizer tubing was on the floor next to recliner.
On 03/07/2023 at 1:56 p.m., Resident 21'a nebulizer tubing and equipment including mouthpiece was observed to be out of the bag sitting next to machine. There was a clear liquid noted to be in the chamber of the nebulizer equipment. On the floor next to the resident's recliner was an empty clear plastic bag that was unlabeled and not dated. A second clear plastic bag was observed to be on the window ledge that contained the word Nebulizer on the bag with the resident's name on it and it was dated 12/4/2022.
On 03/08/23 at 1:40 p.m., Resident 21's nebulizer tubing and equipment was observed to be inside a clear plastic bag that was unlabeled and not dated hanging on the nebulizer machine. No date noted to be on the nebulizer tubing. A clear liquid was noted to be inside the chamber of the nebulizer equipment.
1b. During an observation on 3/06/23 at 10:25 a.m., Resident 21 was observed sitting in her recliner receiving supplemental oxygen per a nasal cannula from an oxygen concentrator on the floor. A nebulizer machine was observed on a table next to the recliner, the mouthpiece and tubing were observed to be out of a bag next to the machine.
On 3/8/23 at 11:48 a.m., Licensed Practical Nurse (LPN) 5 was observed entering the resident room to administer a nebulizer aerosol treatment to Resident 21. The nurse removed a mouthpiece from a bag hanging from the side of a bedside stand, poured medication into the nebulizer cup, turned on the nebulizer machine, and stood with the resident as she received her treatment. The nurse was then observed to turn off the nebulizer machine and place the mouthpiece back into the bag hanging from the bedside stand. The nurse was not observed to wash or rinse the remaining medication from the mouthpiece. A second observation of Resident 21's nebulizer mouthpiece on 3/08/23 at 1:42 p.m., indicated small amounts of liquid medication remained in the nebulizer cup.
On 3/09/23 at 8:46 a.m., Resident 21's nebulizer mouthpiece and tubing were observed to be inside a clear plastic bag hanging from the bedside stand, the mouthpiece was observed to still contain liquid inside the nebulizer cup.
On 3/09/23 at 11:44 a.m., Resident 21 was observed sitting in a wheelchair in her room after LPN 6 was observed giving her a nebulizer treatment. The nebulizer mouthpiece was observed inside a clear plastic bag on the side of a bedside stand, small amounts of liquid medication remained in the nebulizer cup. Resident 21 indicated staff changed the mouthpiece weekly, and the resident cleaned off the outside of the mouthpiece with alcohol preps. To her knowledge, no one took the mouthpiece apart and rinsed out the excess medication from the nebulizer cup after each treatment.
Resident 21's record was reviewed on 3/09/23 at 8:58 a.m. Diagnoses on Resident 21's profile included, but were not limited to, chronic COPD (chronic obstructive pulmonary disease - a group of lung diseases that block airflow and make it difficult to breath) and chronic respiratory failure (develops when the lungs can't get enough oxygen into the blood).
Physician's orders for Resident 21 indicated,
a. On 8/9/22 change nebulizer tubing/set on Sunday 6:00 p.m. - 6:00 a.m.
b. On 1/19/23 ipratropium-albuterol solution (bronchodilators used to treat and prevent wheezing and shortness of breath) for nebulization 0.5 mg (milligrams)/3 mg) per 3 ml (milliliter) give 1 vial three times a day at 6:00 a.m., 12:00 p.m., and 6:00 p.m. and as needed.
A quarterly MDS (Minimum Data Set) assessment (part of the federally mandated process for clinical assessment of all residents in Medicare and Medicaid certified nursing homes), completed on 2/5/23, assessed the resident as receiving oxygen therapy.
A care plans for Resident 21 indicated she was at risk for impaired gas exchange related to COPD with exacerbation with shortness of breath while lying flat, patient to have head of bed elevated. The goal was for the resident to have adequate respiratory functions as evidenced by decreased or absence of dyspnea (shortness of breath), improved breath sounds, decreased or absence of shortness of breath, and improved oximetry (non-invasive method for monitoring oxygen saturation) results. Approaches included oxygen at 3 L (liters), elevate head of bed for shortness of breath when lying flat, monitor the resident, and administer medications as ordered.
An In-Service Class Attendance Record, dated 9/12/22 at 2:00 p.m., subject(s) oxygen application, tubing in bags, and observations, indicated 13 staff members signed as having received the education.
On 3/10/23 at 1:50 p.m., the Executive Director (ED) provided a Respiratory Care: Competency Assessment Form: Aerosolized Medication Nebulizer Treatment, undated, and indicated the process was the one currently being used by the facility. The competency indicated, Verify physician's order. Obtain all necessary equipment and supplies .Attach one end of the tubing to the nebulizer and the other end to the compressor or oxygen source. Remove the nebulizer top, place prescribed medication into the cup, then secure the top back onto the nebulizer. Attach mouthpiece, aerosol mask, or trach mask to the nebulizer [use the reservoir tubing on the t-piece when using a mouthpiece for better medication delivery] .Once the treatment is complete, place the nebulizer in a plastic bag with the resident's name and date. Change out nebulizer weekly per facility policy .
How to Clean a Nebulizer - American Lung Association (11/17/22) at www.lung.org was retrieved from the American Lung Association website. The guidance indicated cleaning a nebulizer was important to prevent the spread of germs and keep individuals from getting sick. It was recommended to wash the parts of the nebulizer after each use, including the mouthpiece or mask, top piece, and medicine cup per manufacturer's instructions. After rinsing, the mouthpiece or mask, top piece, and medication cup were to be left to air-dry in a cool, dry place.